Hamamatsu Deuterium Lamp L2D2 P/N 0035

Deuterium lamp (L2D2 lamp) L7296-50 , Spectral distribution : 160 nm to 400 nm, Aperture diameter : 0.5 mm, Guaranteed life : 2000 h

Electrode Structure Standard
Window Material Synthetic Silica
Spectral Distribution (Short) 160 nm
Spectral Distribution (Long) 400 nm
Aperture Diameter *a 0.5 mm
Output Stability Drift Max. +/- 0.3 %/h
Output Stability Fluctuation (p-p) Typ. 0.005 %
Guaranteed Life *d 2000 h
Required Discharge Starting Voltage Min. 400.0 V dc

Deuterium lamp (L2D2 lamp) L6565 , Spectral distribution : 185 nm to 400 nm, Aperture diameter : 0.5 mm, Guaranteed life : 2000 h

Electrode Structure Standard
Window Material UV Glass
Spectral Distribution (Short) 185 nm
Spectral Distribution (Long) 400 nm
Aperture Diameter *a 1 mm
Output Stability Drift Max. +/- 0.3 %/h
Output Stability Fluctuation (p-p) Typ. 0.005 %
Guaranteed Life *d 4000 h
Required Discharge Starting Voltage Min. 350.0 V dc
Anode Current 300 +/- 30 mA dc
Tube Drop Voltage Typ. 80.0 V dc
Filament Warm-up Voltage 2.5 +/- 0.25 V dc, V ac
Filament Warm-up Current Typ. 4 A dc, A ac
Filament Warm-up Time 20 s
Filament Operating Voltage 1.0 +/- 0.1 V dc
Filament Operating Current Typ. 1.8 A dc

Other Deuterium lamps, Next-generation deuterium lamps that deliver the high stability, high output, and long service life required for spectroscopy.